PULSE 10.3 Firmware Update

Under certain circumstances when using a PULSE system running version 10.3 an erroneous internal calibration error appears. This happens during the initial settling time on specific IDAe acquisition hardware. Until now, this error has not been reported back to the PC software and so has gone undetected.

The conditions that give the error are quite definite and are related to an explicit software build used in conjunction with certain hardware modules.

This patch updates the dynamic firmware used by your acquisition modules and ensures that even if the specific state is achieved, there will be no internal calibration error.

> Download update (261 KB)