MODELO 4394
Acelerómetro CCLD piezoeléctrico en miniatura
Acelerómetro CCLD (Base aislada, cable incluido) diseñado para mediciones en estructuras ligeras, en presencia de vibraciones relativamente elevadas y de alta frecuencia.
Solicitar ofertaType 4394 is a piezoelectric DeltaShear, Unigain accelerometer with side connectors. It features an M3 connection and can be mounted on the test object with an M3 threaded stud, and has a ceramic isolated base.
CHARACTERISTICS
This miniature accelerometer may be treated as a voltage source. The sensitivity is expressed in terms of voltage per unit acceleration (mV/ms-2)
The sensitivity given in the calibration chart is measured at 159.2 Hz. For a 95% confidence level, the accuracy of the factory calibrated unit is +/- 0.7%.
USE SCENARIOS
- Shock and vibration measurement and analysis
- Vibration monitoring
- Modal and structural analysis
- Vibration test control
- Production and quality control
The DeltaShear design involves three piezoelectric elements and three masses in a triangular configuration around a central triangular post.
A clamping ring pre-stresses these elements to give a higher degree of linearity. The signal is collected between the housing and the clamping ring and amplified in the built-in CCLD preamplifier.
*CCLD: Constant Current Line Drive - also known as DeltaTron (IEPE compatible)
TYPE 4394 ACCELEROMETER INCLUDES
- Carrying box
- Calibration chart
- Cable: M3 (M) to 10 - 32 UNF (M), 1,2 m, 250° C / 248° F
- Three steel stud with flange, M3/M3, 3.5 mm (YS-8321)
- Freq. range
- 1 – 25000 Hz
- Temperature
- -50 – 125 ℃
- Weight
- 2.9 gram
- Sensitivity
- 1 mV/ms^-2
- Residual Noise Level in Spec Freq Range (rms) ±
- 2500 µg
- Maximum Operational Level (peak)
- 500 g
- Connector
- M3
- Mounting
- Stud
- Accessory Included
- AO 1381
- Clip/Stud/Screw included
- YS-8381
- Output
- CCLD
- Unigain
- No
- Triaxial
- No
- TEDS
- No
- Applications
- Miniature
- Unit
- mV
- Resonance frequency
- 52 kHz
- Maximum Shock Level (± peak)
- 10000 g